Air Quality Monitoring

Horiba ZNV-7 Zero Air Generator

HORIBA’s APMC-370 is equipped with digital mas flow controllers for calibrating ambient air monitoring analysers.

SKU
N.A.
Price

We will finalise a price once we receive your enquiry.

  • Description

  • Features & Applications

  • Downloads

Zero Gas Generator ZNV-7 has been developed to check the zero point and dilute the span gas sources in order to carry out the calibration of ambient air monitoring systems. The increasing sensitivity of ambient air analysers is required by strict official guidelines, which demands pure and dry air in a sufficient quantity at constant pressure.

  • Ambient air will be sucked through a filter by using an oil-free piston compressor and will be compressed to max. 7 bar.
  • The compressed air will be piped to a cooling spiral followed by coaling filter for eliminating the humidity.
  • The dew point of the gas will be reduced by 35°K using a membrane dryer.
  • The installed container with 7l size will reduce the operation hours to a minimum.
  • The pressure at the outlet can be adjusted in a range of 1 – 4 bar.
  • For cleaning the gas will pass filter cartridges and the Purafill filters after pressure regulating.